缪德仁, 刘玉龙*, 张晓然, 陈宏坤. 多环芳烃类标准物质使用时效性评价[J]. 岩矿测试, 2010, 29(2): 109-112.
引用本文: 缪德仁, 刘玉龙*, 张晓然, 陈宏坤. 多环芳烃类标准物质使用时效性评价[J]. 岩矿测试, 2010, 29(2): 109-112.
MIAO Deren, LIU Yulong, ZHANG Xiaoran, CHEN Hongkun. Timeliness Evaluation for the Application of Certified Reference Materials in Analysis of Polycyclic Aromatic Hydrocarbons[J]. Rock and Mineral Analysis, 2010, 29(2): 109-112.
Citation: MIAO Deren, LIU Yulong, ZHANG Xiaoran, CHEN Hongkun. Timeliness Evaluation for the Application of Certified Reference Materials in Analysis of Polycyclic Aromatic Hydrocarbons[J]. Rock and Mineral Analysis, 2010, 29(2): 109-112.

多环芳烃类标准物质使用时效性评价

Timeliness Evaluation for the Application of Certified Reference Materials in Analysis of Polycyclic Aromatic Hydrocarbons

  • 摘要: 为对半挥发性有机化合物标准物质的正确使用提供指导,选用美国ChemSevice公司生产的PPH-10RPM“甲醇中16种多环芳烃混合标准物质”,将其分装于2 mL聚四氟乙烯衬盖棕色螺口样品瓶中于-18℃避光贮存,采用气相色谱-火焰离子化检测器方法测试各个特性量值随贮存时间的变化。结果表明,标准物质各个特性量测量的标准偏差在0.3~4.0 μg/mL,最大相对标准偏差为3.92%。在置信度为95%时,χ2-检验表明测试结果的精密度适当,评价的测试方法可行;标准物质各特性量值在(100.0±7.2)μg/mL时,测量的偏差不显著。分装后的标准物质至多可保存80天,具体的保存期限还应结合标准物质所剩的有效期确定。

     

    Abstract: In order to offer guidance on correctly using certified reference materials (CRMs) of 16 polycyclic aromatic compounds (PAHs) in methanol (named PPH10RPM) manufactured by ChemService Co. Ltd in application of PAHs analysis, stability of CRMs in opened stocks was tested by gas chromatography with a flame ionization detector (GC/FID). The CRM was transferred into 2 mL brown bottles equipped with Teflonlined screwcaps, stored with minimal headspace and protected from light at -18℃ for 123 days. The variation of property values of the CRMs with storage time was determined. The results showed that the standard deviations of all the property values ranged from 0.3~4.0 μg/mL and the maximum relative standard deviation was 3.92%. In the 95% confidence level, the chisquared test (χ2-test) showed that precision of the measurements was acceptable and indicated the evaluation method was feasible and reliable. In addition, the measurement deviation was not significant when the range of property values of the CRMs were within (100.0±7.2) μg/mL. The experiment data also showed that the preservation period of the CRMs for opened stocks was up to 80 days and the period of validity of the CRMs should be considered.

     

/