汪君, 王頔, 邓长生, 张建梅, 羊鸫, 王斌堂, 朱素霞. 电感耦合等离子体发射光谱法测定地球化学样品中的钍[J]. 岩矿测试, 2014, 33(4): 501-505.
引用本文: 汪君, 王頔, 邓长生, 张建梅, 羊鸫, 王斌堂, 朱素霞. 电感耦合等离子体发射光谱法测定地球化学样品中的钍[J]. 岩矿测试, 2014, 33(4): 501-505.
Jun WANG, Di WANG, Chang-sheng DENG, Jian-mei ZHANG, Dong YANG, Bin-tang WANG, Su-xia ZHU. Determination of Thorium in Geochemical Samples by Inductively Coupled Plasma-Atomic Emission Spectrometry[J]. Rock and Mineral Analysis, 2014, 33(4): 501-505.
Citation: Jun WANG, Di WANG, Chang-sheng DENG, Jian-mei ZHANG, Dong YANG, Bin-tang WANG, Su-xia ZHU. Determination of Thorium in Geochemical Samples by Inductively Coupled Plasma-Atomic Emission Spectrometry[J]. Rock and Mineral Analysis, 2014, 33(4): 501-505.

电感耦合等离子体发射光谱法测定地球化学样品中的钍

Determination of Thorium in Geochemical Samples by Inductively Coupled Plasma-Atomic Emission Spectrometry

  • 摘要: 运用电感耦合等离子体发射光谱(ICP-AES)测定地球化学样品中的钍,样品处理方法通常是采用四酸溶矿。但在实际测定中发现,当钍含量接近本底时,检测结果不稳定;当基体干扰大时,钍含量甚至无法检出。本文对样品前处理方法进行改进,采用过氧化钠碱熔样品,10 g/L氢氧化钠溶液过滤洗涤,再用热的40%盐酸溶解沉淀,ICP-AES法测定钍的含量。通过实验确定了钍的最佳分析谱线为401.913 nm,方法检出限为0.21 μg/g,精密度(RSD,n=6)为7.7%~15.9%,准确度(n=6)为7.0%~10.0%,加标回收率为92.0%~104.0%。经国家标准物质验证,本方法可准确测定钍含量大于0.21 μg/g的样品。方法简便快捷,干扰少,较一般的酸溶ICP-AES测定方法的检出限(0.6~0.7 μg/g)低,适用于大批量实际样品的快速检测。

     

    Abstract: Thorium in geological samples, digested by mixed acid system of HNO3-HCl-HClO4-HF, is measured by using Inductively Coupled Plasma-Atomic Emission Spectrometry (ICP-AES). In the practical analysis, the stability of the results is poor when the content of thorium, moreover, the content of thorium cannot be detected with high matrix effect. In this work, a new method is described for the determination of thorium using ICP-AES. The pretreatment process is optimized by digestion with Na2O2, filtration separated with 10 g/L NaOH and resolved with 40% hot HCl acid. The best analytical line of thorium is 401.913 nm by studying the working conditions of ICP-AES. The method detection limit is 0.21 μg/g, the relative standard deviation (n=7) are from 7.7% to 15.9%, the relative error (n=6) are from 7.0% to 10.0%, and the standard addition recovery are from 92.0% to 104.0%. Furthermore, this method can be used to determine accurately the content of thorium, which is higher than 0.21 μg/g based on verification of certified reference materials. Compared with conventional methods, this method is simple, fast, and highly accurate and has a lower detection limit. It is suitable for practical determination of a large amount of samples.

     

/